Cogprints

Sensors for Rate Responsive Pacing

Dell’Orto, Simonetta and Valli, Paolo and Greco, Enrico Maria (2004) Sensors for Rate Responsive Pacing. [Journal (Paginated)]

Full text available as:

[img] HTML
85Kb
[img]
Preview
PDF
149Kb

Abstract

Advances in pacemaker technology in the 1980s have generated a wide variety of complex multiprogrammable pacemakers and pacing modes. The aim of the present review is to address the different rate responsive pacing modalities presently available in respect to physiological situations and pathological conditions. Rate adaptive pacing has been shown to improve exercise capacity in patients with chronotropic incompetence. A number of activity and metabolic sensors have been proposed and used for rate control. However, all sensors used to optimize pacing rate metabolic demands show typical limitations. To overcome these weaknesses the use of two sensors has been proposed. Indeed an unspecific but fast reacting sensor is combined with a more specific but slower metabolic one. Clinical studies have demonstrated that this methodology is suitable to reproduce normal sinus behavior during different types and loads of exercise. Sensor combinations require adequate sensor blending and cross checking possibly controlled by automatic algorithms for sensors optimization and simplicity of programming. Assessment and possibly deactivation of some automatic functions should be also possible to maximize benefits from the dual sensor system in particular conditions. This is of special relevance in patient whose myocardial contractility is limited such as in subjects with implantable defibrillators and biventricular pacemakers. The concept of closed loop pacing, implementing a negative feedback relating pacing rate and the control signal, will provide new opportunities to optimize dual-sensors system and deserves further investigation. The integration of rate adaptive pacing into defibrillators is the natural consequence of technical evolution.

Item Type:Journal (Paginated)
Keywords:rate responsive pacing, activity sensors, metabolic sensors, physical activity, biventricular pacing, implantable cardioverter defibrillator
Subjects:JOURNALS > Indian Pacing and Electrophysiology Journal
ID Code:4194
Deposited By: Indian Pacing and Electrophysiology, Journal
Deposited On:12 Apr 2005
Last Modified:11 Mar 2011 08:55

References in Article

Select the SEEK icon to attempt to find the referenced article. If it does not appear to be in cogprints you will be forwarded to the paracite service. Poorly formated references will probably not work.

1. Fananapazir L, Bennett DH, Monks P. Contribution of the chronotropic responce to improved exercise performance. Pace 1983; 6:601-608.

2. Rossi P, Rognoni G, Ochetta E, et al. Respiration-dependent ventricular pacing compared with fixed ventricular and atrial-ventricular synchronous pacing. J Am Coll Cardiol 1985; 6:646-652.

3. Furman S. Rate-modulated pacing. Circulation 1990; 82:1081-1094.

4. Ovsyshcher I, Guetta V, Bondy C, et al. First derivative of right ventricular pressure, dP/dt, as a sensor for a rate adaptive VVI pacemaker: initial experience. Pace 1991; 15:211-218.

5. Lau C, Butrous G, Ward D, et al. Comparison of exercise performance of six rate-adaptive right ventricular cardiac pacemakers. Am J Cardiol 1991; 63:833-838.

6. Lau C, Antoniou A, Ward DE, et al. Initial clinical experience with minute ventilation sensing rate modulated pacemaker: improvement in exercise capacity and symptomatology. Pace 1988; 11:1815-1822.

7. Benditt D, Mianulli M, Fetter J, et al. Single-chamber cardiac pacing with activity initiated chronotropic response. Evaluation by cardiopulmonary exercise test. Circulation 1987; 75:184-191.

8. Alt E. What is the ideal rate-adaptive sensor for patients with impiantable cardioverter defibrillators: lessons from cardiac pacing. Am J Cardiol 1999; 83:17-23.

9. Daubert C, Mabo P, Pouillot C, Lelong B. Atrial chronotropic incompetence: implications for DDDR pacing. In: Mugica J, editor. NY: 1991: 251-271.

10. Powers SK, Howley ET. Exercise physiology, theory and application to fitness and performance. Brown,C., 1990.

11. Howley ET, Don Franks B. In: Champain, editor. Health fitness-instructors' handbook. 1992: 42.

12. Lanzberg JS, Franklin JO, Mahavar SK, et al. Benefits of physiologic atrioventricular synchronization for pacing with an exercise rate response. Am J Cardiol 1990; 66:193-197.

13. Iwase M, Hatano K, Saito F, et al. Evaluation by exercise Doppler echocardiography of manteinance of cardiac output during ventricular pacing with or without chronotropic responce. Am J Cardiol 1989; 63:934-938.

14. Fromer J, Kappenberger L, Bottai I. Subjective and objective response to single versus dual chamber pacing. J Electrophysiol 1987; 1:343-349.

15. Buckingham TA, Woodruff RC, Pennington DC, et al. Effect of ventricular function on the exercise hemodynamics of variable rate pacing. J Am Coll Cardiol 1988; 11:1269-1277.

16. Sedney M, Weijers E, Vandervall EE, et al. Short-term and long-term changes of left ventricular volumes during rate-adaptive and single-rate pacing. Pace 1989; 12:1863-1868.

17. Pasquier J, Adamec R, Velebit V, von Segesser L. Long –term improvement of exercise performance in patients with activity-dependent pacemakers. Schweiz Med Wochenschr 1986 Nov 15; 116(46): 1604-7.

18. Pehrsson S, Hjemdhal P, Nordlander R. A comparison of sympathoadrenal activity and cardiac performance at rest and during exercise in patientswith ventricular demand or atrial synchronous pacing. Br Heart J 1988; 60:212-220.

19. Hedman A, Hjemdahl P, Nordlander R, et al. Effects of mental and physical stress on central emodynamics and cardiac sympathetic nerve activity during QT interval sensing rate responsive and fixed-rate ventricular inhibited pacing. Eur Heart J 1990; 11:903-915.

20. Sulke N, Chambers J, Dritsas.A., et al. A randomized double-blind crossover comparison of four rate-responsive pacing modes. J Am Coll Cardiol 1991; 17:696-706.

21. Alagona P Jr. Advances in pacing for the patient with sick sinus syndrome. Curr Opin Cardiol. 1997 Jan; 12(1):3-11.

22. Kato R, Terasawa T, Gotoh T, et al. Antiarrhytmic efficacy of atrial demand (AAI) and rate-responcive atrial pacing. Progress in Clinical Pacing 1988;15-24.

23. Marinoni GP, Gundari F, Orlando F, et al. Can the DDDR mode improve the cardiac performance and have an antiarrhrytmic effect in patients with a chronotropic failure. Eur J Cardiac Pacing Electrophysiol 2, 51. 1992. (Abstract)

24. Adornato E, Monea P, Pennisi V, et al. Comparative evaluation of DDDR and DDD pacing modes in the prevention of atrial fibrillation. Pace 16, 1145. 1993. (Abstract)

25. Wirtfiel A, Goedel Meinem L, Bock T, et al. Central venous oxygen saturation for the control of automatic rate responsive pacing. Pace 1982; 5:829.

26. Camilli L, Alcidi L, Shaoland E, et al. Results, problems, perspectives with the autoregulating pacemaker. Pace 1983; 6:488.

27. Rickards AF, Donaldson R, Thalen H. The use of QT interval to determin pacing rate. early clinical experience. Pace 1983; 6:346.

28. Hummen DP, Anderson K, Brumwell D, et al. A pacemaker which automatically increases its rate with physical activity. Cardiac Pacing: Proceeding of the VIIth World Symposium on Cardiac Pacing 1983;259.

29. Rossi P, Aina F, Rognoni G, et al. Increasing cardiac rate by tracking the respiratory rate. Pace 1984; 7:1246-1256.

30. Salo RW, Pederson B, Olive AL, et al. Continuous ventricular volume assessement for diagnosis and pacemaker control. Pace 1984; 7:1267.

31. Fearnot N, Smith HJ, Geddes LA. A review of pacemakers which physiologically increases rate:The DDD rate responcive pacemakers. Prog Cardiovasc Dis 1986; 29:145.

32. Alt E, Volker R, Hogl B, MacCarter D. First clinical results with a new temperature-controlled rate-responsive pacemaker. Comparison of Activitrax and Nova MR pacemakers with VVI/AAI pacing. Circulation. 1988 Nov; 78 (5 Pt 2): III 1 16-24.

33. Griffin J, Jutzy KR, Claude JP, et al. Central body temperatures as a guide for optimal heart rate. Pace 1983; 6:498-501.

34. Mond HL. Rate responsive pacing using a minute ventilation sensor. Pace 1988; 11:1866.

35. Greco EM, Ferrario M, Romano S. Clinical evaluation of Peak Endocardial Acceleration as a sensor for rate responsive pacing. Pace 2003; 26:812-818.

36. Osswald S, Cron T, Graedel C, et al. Closed-loop stimulation using intracardiac impedance as a sensor principle: correlation of right ventricular dP/dT max and intracardiac impedance during dobutamine stress test. Pace 2000; 23:1502-1508.

37. Greco EM, Romano S. Rate responsive pacing: which sensor for which patient? Rhythm control from cardiac evaluation to treatment. Southern Symposium on Cardiac Pacing 2000; 1:277-283.

38. Rossi P. Rate-responsive pacing: biosensor reliability and physiological sensitivity. Pace 1987; 10:454-466.

39. Vai F, Bonnet JL, Ritter P, et al. Relationship between heart rate and minute ventilation, tidal volume and respiratory rate during brief and low level exercise. Pace 1988; 11:1860-1865.

40. Gillette P. Critical analysis of sensors for physiological responsive pacing. Pace 1984; 7:1263-1266.

41. Hauser RG. Techniques for improving cardiac performance with impiantable device. Pace 1984; 7:1234-1239.

42. Southorn PA, Kamat GS, Vasdev GM, Hayes DL. Monitoring equipment induced tachycardia in patients with minute ventilation rate-responsive pacemakers. Br J Anaesth 2000; 84:508-509.

43. Duru F, Radicke D, Wilkoff BL, Cole CL, Adler S, Nelson L, Jensen DN, Strobel U, Portzline G, Candinas R. Influence of posture, breathing pattern, and type of exercise on minute ventilation estimation by pacemaker transthoracic impedance sensor. Pace 2000; 23:1767-1771.

44. Donaldson R, Richards AF. Rate responses pacing using the QT evoked principal. A physiological alternative to atrial synchronous pacemakers. Pace 1983; 6:1344-1349.

45. Rickards AFBT, Corbucci G, et al. An impiantable cardiac accelerometer for monitoring myocardial contractility. Pace 1996; 19:2066-2071.

46. Ritter P, Mugica J. Endocardial acceleration: a new concept for myocardial contractility assessement in impiantable devices. 4th southern symposium on cardiac pacing. How to approch cardiac arrhytmias. 1994: 312-319.

47. Greco EM, Guardini S, Ferrario M, Romano S. How to program rate responsive pacemakers. Pacing Clin Electrophysiol 2000; 23:165-173.

48. Bordachar P, Garrigue S, Reuter S, Hocini M, Kobeissi A, Gaggini G, Jais.P., Haissaguerre M, Clementy J. Hemodynamic assessment of right, left and biventricular pacing by peak endocardial acceleration and echocardiography in patients with end-stage heart failure. Pacing Clin Electrophysiol 2000; 23:1726-1730.

49. Ritter P, et al. Determination of the optimal atrioventricular delay in DDD pacing. Comparison between echo and endocardial peak acceleration measurements. Europace 1999; 1:126-130.

50. Bombardini T, Gaggini G, Marcelli E, Parlapiano M, Plichi G. Peak endocardial acceleration reflects heart contractiliy also in atrial fibrillation. Pace 2000; 23:1381-1385.

51. Clementy J, Barold SS, Garrigue S, Shah DC, Jais P, Le Metayer P, Haissaguerre M. Clinical significance of multiple sensor options: rate response optimization, sensor blending, and trending. Am J Cardiol 1999; 83:166-171.

52. Greco EM, Guardini S, Citelli L. Cardiac rehabilitation in patients with rate responsive pacemakers. Pace 1998; 21:568-575.

Metadata

Repository Staff Only: item control page